FDAClass IIJune 20, 2018

Biomet Modular Primary Tibial Tray Implants recall

Biomet Modular Primary Tibial Tray Implants; Modular Interlok Primary Tray (size 75 mm); Item No. 141215; UDI: (01) 00880304005310 (17) 280306 (10)…

Why was Biomet Modular Primary Tibial Tray Implants recalled?

One lot of 75mm tibial tray is potentially etched and labeled as a 79 mm tibial tray. This may lead to delay in surgery, or immediate revision surgery to remove an incorrectly sized tibial component after it has been cemented.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot 385340 UDI: (01) 00880304005310 (17) 280306 (10) 385340

Product
Biomet Modular Primary Tibial Tray Implants; Modular Interlok Primary Tray (size 75 mm); Item No. 141215; UDI: (01) 00880304005310 (17) 280306 (10)…
Recalling firm
Zimmer Biomet, Inc.
Quantity
16
Distribution
US Distribution to the states of : FL, GA, IN, MA, NM, UT, and WI. Foreign distribution to Canada and Mexico
Recall date
June 20, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls