Outlook ES pump is intended for use with B. Braun recall
Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when…
Why was Outlook ES pump is intended for use with B. Braun recalled?
B. Braun Medical Inc. is informing our customers of a field correction previously performed to address the loss of the drug library in the Outlook Pump Models 620-100 and 620-200. This could have occurred after periods of lost DC (battery) power or out of tolerance low power. This could have led to a delay in therapy, or to delivery of an over or under-dose of a therapy if the user was to rely on
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Models 620-100 and 620-200; multiple serial numbers
- Product
- Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when…
- Recalling firm
- B Braun Medical, Inc.
- Quantity
- 14,161
- Distribution
- Worldwide Disribution: USA (nationwide) and country of: Canada.
- Recall date
- September 18, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)