Color Gram 2 recall
Color Gram 2 (COLOR GRAM 2 - F), these stains are used to stain bacterial and fungal microorganisms for direct examination, according to the Gram…
Why was Color Gram 2 recalled?
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Color Gram 2 (COLOR GRAM 2 - F) (ref: 55542) Lot number G64508
- Product
- Color Gram 2 (COLOR GRAM 2 - F), these stains are used to stain bacterial and fungal microorganisms for direct examination, according to the Gram…
- Recalling firm
- BioMerieux SA
- Quantity
- 8 boxes COLOR GRAM2 with 4 - 240ML bottles in each
- Distribution
- International distribution.
- Recall date
- June 20, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)