Medtronic CareLink SmartSync Device Manager recall
Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria Amplia Compia application, Software Model Number D00U009; 3. Evera MRI application, Software Model Number D00U010; 3. Visia AF application, Software Model Number D00U011; 5. Viva Brava Evera application, Software Model Number D00U012;
Why was Medtronic CareLink SmartSync Device Manager recalled?
In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
1. Software Model Number, D00U003, GTIN 00763000002039, Updated Software Version 6.5.5; 2. Software Model Number, D00U004, GTIN 00763000002046, Updated Software Version 6.5.5; 3. Software Model Number, D00U005, GTIN 00763000002053, Updated Software Version 9.5.2; 4. Software Model Number, D00U006, GTIN 00763000397852, Updated Software Version 4.4.6; 5. Software Model Number, D00U007, GTIN 00763000397869, Updated Software Version 4.4.6; 6. Software Model Number, D00U008, GTIN 00763000397876, Updated Software Version 3.5.5; 7. Software Model Number, D00U009, GTIN 00763000397883, Updated Software Version 3.4.2; 8. Software Model Number, D00U010, GTIN 00763000397890, Updated Software Version 3.4.2; 9. Software Model Number, D00U011, GTIN 00763000397906, Updated Software Version 3.4.2; 10. Software Model Number, D00U012, GTIN 00763000397913, Updated Software Version 3.4.2; 11. Software Model Number, D00U022, GTIN 00763000544300, Updated Software Version 3.7.5.
- Product
- Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria Amplia Compia application, Software Model Number D00U009; 3. Evera MRI application, Software Model Number D00U010; 3. Visia AF application, Software Model Number D00U011; 5. Viva Brava Evera application, Software Model Number D00U012;
- Recalling firm
- Medtronic, Inc.
- Quantity
- 22091 total devices
- Distribution
- Worldwide - US Nationwide distribution.
- Recall date
- July 30, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)