3dimensions Mammography System recall
3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-INTL3D, 3DM-SYS-INTL3D-MOB…
Why was 3dimensions Mammography System recalled?
Complaints have been received of systems developing loose, missing, or broken internal bolts over time.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
"[Model No.; UDI]: [3DM-SYS-INTL2D;15420045510609], [3DM-SYS-INTL2D-MOB;15420045510593], [3DM-SYS-INTL2D-NS;15420045510586], [3DM-SYS-INTL3D;15420045510579], [3DM-SYS-INTL3D-MOB;15420045510562], [3DM-SYS-INTL3D-NS;15420045510555], [3DM-SYS-STD;15420045510524], [3DM-SYS-STD-MOB;15420045510531], [3DM-SYS-STD-NS;15420045510548]; All Serial No.
- Product
- 3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-INTL3D, 3DM-SYS-INTL3D-MOB…
- Recalling firm
- Hologic, Inc
- Quantity
- 14837 units
- Distribution
- US Nationwide distribution.
- Recall date
- May 20, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)