FDAClass IIMay 31, 2017

SENSATION PLUS¿ 8Fr. 50cc Intra-Aortic Balloon Catheter recall

SENSATION PLUS¿ 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK¿ Product Code/Part Number: 0684-00-0576-01U is a cardiac assist…

Why was SENSATION PLUS¿ 8Fr. 50cc Intra-Aortic Balloon Catheter recalled?

Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing of the SENSATION PLUS 8Fr. 50cc, IABC, failures related to the tip seal leak were identified.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

300004515, 3000043376 3000040272 3000038372 added lot numbers: 3000043374 and 3000045513

Product
SENSATION PLUS¿ 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK¿ Product Code/Part Number: 0684-00-0576-01U is a cardiac assist…
Recalling firm
Datascope Corporation
Quantity
996 UNITS
Distribution
Nationwide
Recall date
May 31, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls