365¿M Lithotripsy Fiber- Reusable recall
365¿M Lithotripsy Fiber- Reusable; Sure F!ex laser fibers are intended for use in laser-based surgical applications, including, but not limited to…
Why was 365¿M Lithotripsy Fiber- Reusable recalled?
validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex'" Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model 8013056; All product manufactured from April 8, 2012 to April 8, 2015
- Product
- 365¿M Lithotripsy Fiber- Reusable; Sure F!ex laser fibers are intended for use in laser-based surgical applications, including, but not limited to…
- Recalling firm
- American Medical Systems Innovation Center - Silicon Valley
- Quantity
- 16, 710 total all affected devices
- Distribution
- Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Israel, Poland, United Kingdom, Chile, Italy, Portugal, United States, Colombia, Korea, Qatar, Uruguay, Costa Rica, Lebanon, Romania, Finland, Malaysia and Saudi Arabia.
- Recall date
- July 29, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)