FDAClass IIJune 26, 2024

Gray Revision Instrument System Accessory Tray/Case recall

Gray Revision Instrument System Accessory Tray/Case, Catalog/Part Number: 6210-9-910

Why was Gray Revision Instrument System Accessory Tray/Case recalled?

The design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Part 6210-9-910, UDI-DI: 07613327217780, Lot#: All lots back to 2006.

Product
Gray Revision Instrument System Accessory Tray/Case, Catalog/Part Number: 6210-9-910
Recalling firm
Howmedica Osteonics Corp.
Quantity
209 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Brazil, Canada, Chile, Denmark, Hong Kong, Iran, Italy, Japan, Kuwait, Mexico, New Zealand, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, UK, Venlo.
Recall date
June 26, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)