FDAClass IIAugust 14, 2019

SenTec Standard Starter Set with and without Service recall

SenTec Standard Starter Set with and without Service Gas (2 changers each) - Product Usage: Is indicated for continuous, non-invasive monitoring of…

Why was SenTec Standard Starter Set with and without Service recalled?

There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Standard Starter Set Lots: 180932, 180959, 181019, 181082, 190005, 190041, 190093, 190221, and 190282. Standard Starter Set (without Service Gas) Lots:180972 and 190078

Product
SenTec Standard Starter Set with and without Service Gas (2 changers each) - Product Usage: Is indicated for continuous, non-invasive monitoring of…
Recalling firm
SenTec AG
Quantity
22 pcs
Distribution
Worldwide distribution - US Nationwide in the state of Missouri and countries of Argentina, Australia, Austria, Belgium, Chile, China, Columbia, Costa Rica, Czech Republic, Denmark, Ecuador, Europe, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Indonesia, Israel, Italy, Kuwait, Malaysia, Finland, Portugal, Romania, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, United Arab Emirates and United Kingdom.
Recall date
August 14, 2019
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls