Opteryx Variable Rescue Screw recall
Opteryx Variable Rescue Screw, Part #CP2214, UDI: *+M711CP22140/$$801716012*. Packaged in heat-sealed plastic bag. Product Usage: 4.5mm x 14mm Anterior Cervical Plate System Variable Rescue Screw. The Opteryx¿ Cervical Plate System is intended as a temporary fixation device used for correction and stabilization of the cervical spine. Variable angle screws are included to allow the physician greater flexibility when placing the plates and securing it to the vertebrae.
Why was Opteryx Variable Rescue Screw recalled?
Opteryx Variable Rescue Screws are engraved with incorrect (larger) size (16 instead of 14).
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot #1601
- Product
- Opteryx Variable Rescue Screw, Part #CP2214, UDI: *+M711CP22140/$$801716012*. Packaged in heat-sealed plastic bag. Product Usage: 4.5mm x 14mm Anterior Cervical Plate System Variable Rescue Screw. The Opteryx¿ Cervical Plate System is intended as a temporary fixation device used for correction and stabilization of the cervical spine. Variable angle screws are included to allow the physician greater flexibility when placing the plates and securing it to the vertebrae.
- Recalling firm
- Innovasis, Inc
- Quantity
- 21
- Distribution
- US Nationwide Distribution in the states of CA, UT, ID, TX, OH, and MN.
- Recall date
- May 31, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)