Knee Products: 141356 Regenerex Series-A Patella 3 Peg recall: Endotoxin
Knee Products: 141356 Regenerex Series-A Patella 3 Peg, 31 MM 141358 Regenerex Series-A Patella 3 Peg, 37 MM Product Usage: Knee prosthesis
Why was Knee Products: 141356 Regenerex Series-A Patella 3 Peg recalled?
Potential presence of elevated endotoxin levels that exceed the specification limit
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Numbers: UDI Number 1) 561760 (01)00880304554412(17)211229(10)561760; 2) 091090 (01)00880304506084(17)211214(10)091090
- Product
- Knee Products: 141356 Regenerex Series-A Patella 3 Peg, 31 MM 141358 Regenerex Series-A Patella 3 Peg, 37 MM Product Usage: Knee prosthesis
- Recalling firm
- Biomet, Inc.
- Quantity
- 15 units
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Canada, AHMADABAD, GUJARAT, BANGKOK, BELROSE NSW, BUENOS AIRES, CHILE, COSTA RICA, HIRATSUKA CITY, HUECHURABA SANTIAGO, INDIA, JAPAN, MEXICO DF, NETHERLANDS, PICHINCHA, EC, SAN JOSE, SINGAPORE, TAIWAN, THAILAND, YONGSAN-GU SEOUL.
- Recall date
- June 3, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)