FDAClass IAugust 15, 2012

AUTOMIX 3+3 Compounder System recall

AUTOMIX 3+3 Compounder System, product 2M8286 (115V) and 2M8286K (230V), automated nutrition compounders; Product Usage: The Automix Compounder…

Why was AUTOMIX 3+3 Compounder System recalled?

After fluid ingress into the keypad of the AUTOMIX Compounder's (AUTOMIX) control module, the AUTOMIX may generate an incorrect device response to an AUTOMIX operator's key press. Also, the AUTOMIX system may experience an intermittent electrical failure, in which unexpected pump motor operations will occur. Also, the AUTOMIX system may intermittently experience an electrical failure where the compounder may stop compounding without an alarm and not be detected by the user.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Product codes 2M8286 and 2M8286K, all serial numbers

Product
AUTOMIX 3+3 Compounder System, product 2M8286 (115V) and 2M8286K (230V), automated nutrition compounders; Product Usage: The Automix Compounder…
Recalling firm
Baxter Healthcare Corp.
Quantity
3106 units
Distribution
Worldwide Distribution - USA - Nationwide including Puerto Rico, and internationally to Brazil, Canada, Colombia, Costa Rica, Czech Republic, El Salvador, Egypt, Germany, Guatemala, Honduras, Luxembourg, Mexico, Panama, Saudi Arabia, Singapore, Switzerland, Taiwan, Trinidad, Turkey, United Arab Emirates and United Kingdom.
Recall date
August 15, 2012
Classification
Class I
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls