VITROS 4600 Chemistry System recall: Software Issue
VITROS 4600 Chemistry System (VITROS 5,1 FS System family member) --- The VITROS 4600 Chemistry System is intended for use in the in vitro…
Why was VITROS 4600 Chemistry System recalled?
Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample identification (ID) number.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial Numbers J46000108 - J46000440 (J Numbers are analogous to Serial Numbers, for VITROS 4600 systems the J Numbers start with 4600XXXX)
- Product
- VITROS 4600 Chemistry System (VITROS 5,1 FS System family member) --- The VITROS 4600 Chemistry System is intended for use in the in vitro…
- Recalling firm
- Ortho-Clinical Diagnostics
- Quantity
- Domestic: 81, Foreign: 238
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
- Recall date
- August 13, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)