Aeos Robotic Digital Microscope recall
Aeos Robotic Digital Microscope, Product Code: PV010
Why was Aeos Robotic Digital Microscope recalled?
Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Product Code: PV010, UDI/DI: 04046955206742; Serial Numbers: 1011, 1005, 1057, 1032, 1038, 1054, 1036, 1060, 1056, 1089, 1093, 1027, 1061, 1004, 1026, 1035, 1097, 1084, 1063
- Product
- Aeos Robotic Digital Microscope, Product Code: PV010
- Recalling firm
- Aesculap Inc
- Quantity
- 19 units
- Distribution
- US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.
- Recall date
- June 19, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)