I-STAT EG7+ Cartridge Abbott Point of Care recall
I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium…
Why was I-STAT EG7+ Cartridge Abbott Point of Care recalled?
i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
List number 03P76-25 Lot number S10246A
- Product
- I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium…
- Recalling firm
- Abbott Point Of Care Inc.
- Quantity
- 52,750 cartridges
- Distribution
- USA ( nationwide )
- Recall date
- August 8, 2012
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)