FDAClass IIAugust 6, 2014

FlowCOUPLER. An implantable device that is used to recall

FlowCOUPLER. An implantable device that is used to detect blood flow in vessels. Used in the end-to-end anastomosis of veins and arteries normally…

Why was FlowCOUPLER. An implantable device that is used to recalled?

Baxter Healthcare is recalling the FlowCOUPLER Device because there is a possibility that some units may not maintain electrical continuity during handling intra-operatively and post-operatively, resulting in a loss of Doppler signal. This may result in the inability to monitor anastomotic patency, with the possibility that blood flow may not be appropriately monitored.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Number SPSGT14-02B0027a SPSGT14-04D0005 SPSGT14-04D0023 SPSGT13-10J0016 SPSGT13-11K0041 SPSGT13-12L0008 SPSGT14-01A0009 SPSGT14-02B0026 SPSGT14-04D0015 SPSGT14-04D0026 SPSGT13-10J0017 SPSGT13-11K0039 SPSGT13-11K0042 SPSGT13-12L0015 SPSGT14-01A0028 SPSGT14-02B0028 SPSGT14-02B0033 SPSGT14-04D0014 SPSGT14-04D0027 SPSGT14-05E0005 SPSGT13-10J0018 SPSGT13-11K0040 SPSGT13-12L0007 SPSGT13-12L0012 SPSGT14-01A0027 Part No. 5151-01200-010 5151-01200-010 5151-01200-010 5151-01200-011 5151-01200-011 5151-01200-011 5151-01200-011 5151-01250-010 5151-01250-010 5151-01250-010 5151-01250-011 5151-01250-011 5151-01250-011 5151-01250-011 5151-01250-011 5151-01300-010 5151-01300-010 5151-01300-010 5151-01300-010 5151-01300-010 5151-01300-011 5151-01300-011 5151-01300-011 5151-01300-011 5151-01300-011

Product
FlowCOUPLER. An implantable device that is used to detect blood flow in vessels. Used in the end-to-end anastomosis of veins and arteries normally…
Recalling firm
Baxter Healthcare Corporation
Quantity
1,315 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of United Kingdom, Germany, Sweden, Spain, Israel and Denmark.
Recall date
August 6, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)