FDAClass IIJune 13, 2018

Locking Screw recall

Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) D 3.2MM X 10MM (002-E-32010) b) D 3.2MM X 12MM (002-E-32012)…

Why was Locking Screw recalled?

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

(UDI)/Lot: a) (18056099646505) B1075433; b) (18056099646512) B1075434; c) (18056099646529) B1075844; d) (18056099646536) B1075432; e) (18056099646543) B1075436; f) (18056099646550) B1075826; g) (18056099646567) B1075837; h) (18056099646574) B1084804; i) (18056099646581) B1075835; j) (18056099646598) B1082233; k) (18056099646604) B1204272, B1075829; l) (18056099646611) B1204977, B1082245; m) (18056099646628) B1215471, B1204978, B1082247; n) (18056099646635) B1204979, B1084800; o) (18056099646642) B1204582, B1082248; p) (18056099646659) B1218842, B1204980; q) (18056099646666) B1218840, B1204271, B1089889; r) (18056099646673) B1218843, B1195750, B1083640; s) (18056099646680) B1204981, B1083837; t) (18056099646697) B1204982, B1083846; u) (18056099646703) B1204584, B1083852, B1071600, B1071601

Product
Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) D 3.2MM X 10MM (002-E-32010) b) D 3.2MM X 12MM (002-E-32012)…
Recalling firm
Orthofix, Inc
Quantity
1574
Distribution
U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
Recall date
June 13, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)