VITROS Chemistry Products CHE Slides recall
VITROS Chemistry Products CHE Slides; Catalog Numbers: (1) 191 4605 (300 slides), (2) 800 4707 (90 slides);
Why was VITROS Chemistry Products CHE Slides recalled?
Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog Numbers: (1) 191 4605 (300 slides), (2) 800 4707 (90 slides); UDI-DI: (1) 10758750000241, (2) 10758750004225; Lot Numbers: All lots within expiry.
- Product
- VITROS Chemistry Products CHE Slides; Catalog Numbers: (1) 191 4605 (300 slides), (2) 800 4707 (90 slides);
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Quantity
- 27, 330 US/ 19,679,730 OUS
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.
- Recall date
- July 23, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)