FDAClass IIJuly 23, 2025

Brand Name: VITROS Chemistry Products Product Name recall

Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806 7324 Software Version…

Why was Brand Name: VITROS Chemistry Products Product Name recalled?

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Code: Product Code: 806 7324 (UDI: 10758750004317) Lots currently within expiry: V1665 - Exp.8/14/2025 B2245 - Exp 3/25/2026 Y2110 - Exp. 3/25/2026 E2828 - Exp. 2/3/2027

Product
Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806 7324 Software Version…
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
37429 units (US)/11,703 units (OUS)
Distribution
Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.
Recall date
July 23, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls