Benchmark XT/LT and Discovery XT recall
Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS, N750-BMKLT-FS, N750-DIS-FS, N750-DISXT-FS. Product Usage: Usage: The…
Why was Benchmark XT/LT and Discovery XT recalled?
Ventana Medical System is initiating the recall of the BenchMark and DISCOVERY series instruments because there is a possibility that the heating element could separate from the underside of the heater cap and cause a non-staining or inconsistent staining event. Additionally, the system may not detect this anomaly and therefore it may not generate an error message for the slide position.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
BenchMark XT Instruments-S/N: 711869-712465; BenchMark LT instruments-S/N: 610321-610345. Discovery XT Instruments-S/N: 610321-610345.
- Product
- Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS, N750-BMKLT-FS, N750-DIS-FS, N750-DISXT-FS. Product Usage: Usage: The…
- Recalling firm
- Ventana Medical Systems Inc
- Quantity
- 1217 units total
- Distribution
- Worldwide Distribution - USA Nationwide and the countries of Canada, Europe, Asia, Japan.
- Recall date
- September 4, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)