ARCHITECT iGentamicin Reagent Kit - Product Usage: The recall
ARCHITECT iGentamicin Reagent Kit - Product Usage: The measurements obtained are used in the diagnosis and treatment of gentamicin overdose and in…
Why was ARCHITECT iGentamicin Reagent Kit - Product Usage: The recalled?
Abbott identified that samples tested for ARCHITECT STAT High Sensitive Troponin I and STAT High Sensitivity Troponin I assays (LN 3P25 and LN 2R98) may show interaction when processed directly after the ARCHITECT iGentamicin (LN 1P31) assay and patient results might be impacted.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
MODEL: 1P31-25 LOT NUMBERS: 04281BE00, 08510BE00 and 11221BE00 Shelf life per LOT: 04281BE00;2020-MAY-21 08510BE00; 2020-AUG-08 11221BE00;2020-NOV-18
- Product
- ARCHITECT iGentamicin Reagent Kit - Product Usage: The measurements obtained are used in the diagnosis and treatment of gentamicin overdose and in…
- Recalling firm
- Abbott Gmbh & Co. KG
- Quantity
- 6087 Reagent Kits
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CT, DE, FL, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MT, ND, NE, NJ, NV, NY, OH, OK, PA, RI, SC, TN,TX, UT, VA, VT, WA, WI, PR. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Barbados, Belarus, Belgium, Brazil, Brunei, Canada, Cayman Islands, Chile, China, Costa Rica, Croatia, Curacao, Czech Republic, Faroe Islands, France, Germany, Greece, Greenland, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Latvia, Lebanon, Lithuania, Malaysia, Mauritius, Namibia, Netherlands, New Zealand, Norway, Oman, Palestinian Territory, Philippines, Poland, Portugal, Republic of Yemen, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tunisia, UAE, United Kingdom and Uruguay.
- Recall date
- June 3, 2020
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)