FDAClass IIJune 3, 2020

FlexLab Automation System used with the following recall

FlexLab Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-276, FLX-282, FLX-283), HSQ Interface Module…

Why was FlexLab Automation System used with the following recalled?

The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: FLX.0133 FLX.0191 FLX.0192 FLX.0197 FLX.0198 FLX.0201 FLX.0202 FLX.027 FLX.0200 FLX.0181

Product
FlexLab Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-276, FLX-282, FLX-283), HSQ Interface Module…
Recalling firm
Inpeco S.A.
Quantity
10 systems
Distribution
The products were distributed to the following US States: IL and NY.
Recall date
June 3, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)