FDAClass IIJune 3, 2020

Optilite Optilite Rheumatoid Factor Kit recall

Optilite Optilite Rheumatoid Factor Kit, Ref: LK151.OPT.A, IVD, Rx Only, UDI: 05051700020015 (kit ea. containing four components)

Why was Optilite Optilite Rheumatoid Factor Kit recalled?

Assigned Rheumatoid Factor (RF) calibrator value for calibrator component included in kit used to aid in the diagnosis of rheumatoid arthritis over recovers by 7% in comparison to the reference material RF serum. This could result in reporting abnormal (false positive) results, which may lead to unnecessary additional serological testing and a delay in treatment.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Kit Lot Number: 428224

Product
Optilite Optilite Rheumatoid Factor Kit, Ref: LK151.OPT.A, IVD, Rx Only, UDI: 05051700020015 (kit ea. containing four components)
Recalling firm
The Binding Site Group, Ltd.
Quantity
478 kits
Distribution
US: AL, AR, AZ, FL, GA, KY, MA, MD, MI, MN, MS, NC, NE, NY, OR, PA, TX, VT, WA, WI OUS: None
Recall date
June 3, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls