FDAClass IIJune 3, 2020

ARTIS Pheno - Interventional Fluoroscopic X-Ray System recall

ARTIS Pheno - Interventional Fluoroscopic X-Ray System Model: 10849000 - Product Usage: is a family of dedicated angiography systems developed for…

Why was ARTIS Pheno - Interventional Fluoroscopic X-Ray System recalled?

If the C-arm leaves its intended travel path due to a fault within the drivetrain, movements may be impacted or impossible and the area of interest cannot be reached. In the event the C-arms movements are not possible, the system can only be returned to normal operation with the support of a field service engineer

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: 164011 164013 164018 164019 164021 164022 164024 164025 164026 164027 164028 164033 164034 164048 164049 164053 164054 164055 164057 164060 164061 164063 164064 164066 164068 164069 164071 164072 164073 164074 164075 164080 164083 164084 164086 164088 164094 164095 164097 164098 164100 164105 164113 164115 164121 164124 164125 164132 164134 164138 164140 164142 164152 164153 164155 164157 164158 164159 164160 164161 164163 164168 164171 164172 164173 164183 164184 164186 164187 164188 164189 164190 164196 164201 164202 164204 164205 164210 164211 164212 164213 164216 164227 164228 164231 164234 164240 164241 164242 164244 164250 164253 164257 164263 164267 164269 164270 164271 164275 164277 164279 164282 164287 164292 164293

Product
ARTIS Pheno - Interventional Fluoroscopic X-Ray System Model: 10849000 - Product Usage: is a family of dedicated angiography systems developed for…
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
105 systems
Distribution
US Nationwide distribution.
Recall date
June 3, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls