FDAClass IIJuly 12, 2023

BD Synapsys Informatics Solution recall

BD Synapsys Informatics Solution, Catalog Number 444150

Why was BD Synapsys Informatics Solution recalled?

There is a risk of mis-association of patient demographic data for users who choose to report results manually when their Service Control Unit (SCU) is operating near capacity and are using save actions (Skip, Final Discard, Mark as Read, Save) to process patient results rapidly.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 00382904441500;

Product
BD Synapsys Informatics Solution, Catalog Number 444150
Recalling firm
Becton Dickinson & Co.
Quantity
102 systems
Distribution
US States: CA, CO, IA, IL, MI, MN, MO, NY, OH, OK, SC, TN, TX, UT, WI. Australia, Austria, Belgium, Canada, Denmark, France, Germany, Italy, Japan, Luxembourg, Netherlands, New Zealand, Norway, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom
Recall date
July 12, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)