Proton therapy system -Proteus 235 aka Proteus Plus. A recall
Proton therapy system -Proteus 235 aka Proteus Plus. A medical device designed to produce and deliver a proton beam for the treatment of patients with…
Why was Proton therapy system -Proteus 235 aka Proteus Plus. A recalled?
A PBS (Pencil Beam Scanning) irradiation may pause for different reasons. After a pause, irradiation may restart from the beginning instead of recalculating the field based on the already delivered dose.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial numbers: PAT.003, PAT.006
- Product
- Proton therapy system -Proteus 235 aka Proteus Plus. A medical device designed to produce and deliver a proton beam for the treatment of patients with…
- Recalling firm
- Ion Beam Applications S.A.
- Quantity
- 2
- Distribution
- FL
- Recall date
- May 24, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)