FDAClass IIJuly 29, 2015

Safety PICC Insertion Tray with Chloraprep OneStep recall: Sterility Concern

Safety PICC Insertion Tray with Chloraprep OneStep - Model #5026CL-9 - 12 pr case. Packaged for Physician use only.

Why was Safety PICC Insertion Tray with Chloraprep OneStep recalled?

An INCORRECT item number is packed within the OUTSIDE "PIGGYBACKED" PACKAGE. The interior sealed STERILE KIT is correct and NO adverse issues affect that part of the item. The NON-STERILE OUTSIDE "PIGGYBACKED" PACKAGE mistakenly contains ITEM #8881579121 - Syringe prefilled, Saline Flush, 09% Sodium Chloride and is a NONSTERILE PRODUCT that CANNOT BE DROPPED INTO THE STERILE FIELD.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot#92414 Exp Date 11/1/15, Lot#100214 Exp Date 11/1/15 Lot#100914 Exp Date 11/1/15

Product
Safety PICC Insertion Tray with Chloraprep OneStep - Model #5026CL-9 - 12 pr case. Packaged for Physician use only.
Recalling firm
Custom Healthcare Systems, Inc.
Quantity
3 Cs-Lot#92414 , 25 Cs-Lot#100214, 24 Cs-Lot#100914
Distribution
US Distribution to CT.
Recall date
July 29, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls