FDAClass IIAugust 8, 2012

Nextgen Complete Knee Solution Femoral and Provisional recall

Nextgen Complete Knee Solution Femoral and Provisional Impactor/ Extractor, REF 00-5901-026-00, Non-Sterile Zimmer, Warsaw Indiana The instrument is…

Why was Nextgen Complete Knee Solution Femoral and Provisional recalled?

The design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design. The spring clip may fracture or fall off of the instrument during use, and a device fragment may be left in the surgical site. The risk of disassembly or fracture occurs when the instrument is typicall

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Part 00-5901-026-00 Lot 11002148, 11002149, 11002791, 11002792, 11002831, 11003439, 11003449, 61800676, 61814993, 61909414, 61924920, 61980656, 61985378, 61988272, 61990656, 62003795, 62011220, 77000065, 77000573, 77000574, 77000575, 77000583, 77000654, 77000786, 77000787, 77000788, 77000789, 77000790, 77000795, 77000797, 77000798, 77000799, 77000804, 77000848, 77000849, 77000872, 77000873, 77001039, 77001048, 77001096, 77001097, 77001118, and 77001119

Product
Nextgen Complete Knee Solution Femoral and Provisional Impactor/ Extractor, REF 00-5901-026-00, Non-Sterile Zimmer, Warsaw Indiana The instrument is…
Recalling firm
Zimmer Inc.
Quantity
1662 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Singapore, China, Hong Kong, Canada, Australia, Japan, Europe, Germany, India, and Finland.
Recall date
August 8, 2012
Classification
Class II
Status
Terminated
Agency
FDA (device)