FDAClass IIMay 24, 2017

ADVIA Centaur Systems Insulin ReadyPack recall

ADVIA Centaur Systems Insulin ReadyPack; Test Code: IRI; Catalog Number: 02230141, for in vitro diagnostic use in the quantitative determination of…

Why was ADVIA Centaur Systems Insulin ReadyPack recalled?

Current lots of ADVIA Centaur/XP/XPT/CP Insulin (IRI) assay recover approximately 40% higher than the World Health Organization (WHO) 1st IRP 66/304 standardization based on slope values. Other performance characteristics are being met per the assay Instructions for Use. Overall risk to health is negligible and Siemens is not recommending a review of previously generated results.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Kit Lots Ending: 169, 170, 173, 174

Product
ADVIA Centaur Systems Insulin ReadyPack; Test Code: IRI; Catalog Number: 02230141, for in vitro diagnostic use in the quantitative determination of…
Recalling firm
Siemens Healthcare Diagnostics, Inc
Distribution
US OUS: Afghanistan, Albania, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Latvia, Lithuania, Luxembourg, Macedonia, Malaysia, Malta, Martinique, Mexico, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of Korea, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay, Vatican, Vietnam, Austria, Belgium, Bulgaria, Switzerland, Cyprus, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, Great Britain, Hungary, Italy, Luxembourg, Latvia, Netherlands, Norway, Greece, Croatia, Poland, Portugal, Romania, Sweden, Slovak Republic, Turkey
Recall date
May 24, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls