Tibial Punch Tower Assembly. Non-Sterile. CAT NO. recall
Tibial Punch Tower Assembly. Non-Sterile. CAT NO. 8000-1089. Howmedica Osteonics Corp., 59 Route 17, Allendale, NJ 07401. During primary Scorpio TKA…
Why was Tibial Punch Tower Assembly. Non-Sterile. CAT NO. recalled?
Stryker Orthopaedics received a report indicating that a Tibial Keel Punch could not be inserted into the slot of the Tibial Punch Tower. The slot that guides the stem of the keel punch into the channel of the tower was later determined to be undersized.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog Number 8000-1089. Lot Number(s): RD6E087, RD6E187; RD6E083; RD6E082; RD6C178; RD5W268; RD5W267; RD5V260; RD5W320; RD5V240; RD5V239; RD5V238; RD5N407; RD5S104; RD5S103; RD5T177; RD5T179; RD5S259; RD5T178; RD5N406; RD5S300; RD5T298; RD5W381; RD6E086; RD6E089; RD6E090; RD6E091; RD6E158; RD6E159; RD6E160; RD6E186.
- Product
- Tibial Punch Tower Assembly. Non-Sterile. CAT NO. 8000-1089. Howmedica Osteonics Corp., 59 Route 17, Allendale, NJ 07401. During primary Scorpio TKA…
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Quantity
- 152 units
- Distribution
- Worldwide Distribution
- Recall date
- August 1, 2012
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)