FDAClass IIMay 27, 2020

Dako Autostainer instruments: Autostainer Link 48 recall

Dako Autostainer instruments: Autostainer Link 48, Model AS480; Autostainer Plus Link Instrument, Model AS100; Autostainer, Model S3400; and…

Why was Dako Autostainer instruments: Autostainer Link 48 recalled?

The Autostainer instrument may not have a syringe tray installed which could result in a small volume of excess buffer being applied to several slides. In addition to the syringe tray, the updated Basic User Guide will reduce the potential risk of false negative results in the case of leakage of buffer and/or reagents that may impact staining.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All serial numbers of devices manufactured between 7/22/1997-6/28/2018 that do not have an installed syringe tray, UDI 05700572035497.

Product
Dako Autostainer instruments: Autostainer Link 48, Model AS480; Autostainer Plus Link Instrument, Model AS100; Autostainer, Model S3400; and…
Recalling firm
Dako Denmark A/S
Quantity
796 devices
Distribution
Distribution was nationwide. There was government/military distribution. Foreign distribution was made to Canada, Australia, Bangaladesh, Belarus, Brazil, Bulgaria, Chile, China, Colombia, Croatia, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Macedonia, Morocco, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Poland, Qatar, Romania, Russia, Saudi Arabia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Uruguay, and Vietnam.
Recall date
May 27, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls