FDAClass IAugust 21, 2019

Centurion Convenience Kits recall

Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110931) contained inside kits DYNJ52509A and…

Why was Centurion Convenience Kits recalled?

This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to an issue with the consistency of the inner lumen thickness of the silicone tubing segment which directly impacts amount of occlusion force necessary to fully seal the tube. This defect prevents the pump from complete occlusion and results in the potential for over-infusion of drug product into the patient.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

lots 2018090590 2018112190 2018090690 2018101190 2018101590 2018121190 2019013190 2019030490 2019040190

Product
Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110931) contained inside kits DYNJ52509A and…
Recalling firm
Medline Industries Inc
Quantity
4890
Distribution
One distributor located in Illinois.
Recall date
August 21, 2019
Classification
Class I
Status
Terminated
Agency
FDA (device)