FDAClass IIMay 17, 2017

Savaria SL-1000 Stairlift recall

Savaria SL-1000 Stairlift

Why was Savaria SL-1000 Stairlift recalled?

The electrical controller, combined with motor, causes feedback, damaging a component on the control board, starting a fire.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

s/n or lot number - Job Number lots 671494 - 672668

Product
Savaria SL-1000 Stairlift
Recalling firm
Savaria Concord Lifts, Inc.
Quantity
1425 in the United States  734 in Canada
Distribution
Distribution in the United States and Canada
Recall date
May 17, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)