Zyno Medical Administration Set -105"ADMIN SET 20DP recall
Zyno Medical Administration Set -105"ADMIN SET 20DP W/FILTER 2YSITES CLMP- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code…
Why was Zyno Medical Administration Set -105"ADMIN SET 20DP recalled?
Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Numbers: 18046309, 18026344 UDl/UPC/GTIN: 00814371020204
- Product
- Zyno Medical Administration Set -105"ADMIN SET 20DP W/FILTER 2YSITES CLMP- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code…
- Recalling firm
- Zyno Medical LLC
- Quantity
- 360 cases (50/case)=18,000
- Distribution
- MA
- Recall date
- July 31, 2019
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)