Linear Straight Broach Handle recall
Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZ Clean" Linear Broach Handle Greatbatch…
Why was Linear Straight Broach Handle recalled?
Greatbatch Medical has voluntarily initiated a Recall for the Linear Straight Broach Handle (Rasp Handle) and EZ Clean" Linear Broach Handle. Greatbatch Medical has received reports that that the product is breaking at the welds and that the locating pin at the distal end of the instrument is breaking off during use.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot numbers : Linear Straight Broach Handle (Rasp Handle): 2988946, 3138186, 3185601, 3287315, 3300879, 3333172, 3433392, 3332697. EZ Clean" Linear Broach Handle: 3508491.
- Product
- Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZ Clean" Linear Broach Handle Greatbatch…
- Recalling firm
- Greatbatch Medical
- Quantity
- 205
- Distribution
- US Distribution to the state of : TX
- Recall date
- May 17, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)