B-Hydroxybutyrate LiquiColor recall
B-Hydroxybutyrate LiquiColor, Item code CH2440058 - Product Usage: Product is used for the quantitative determination of B-Hydroxybutyrate in human…
Why was B-Hydroxybutyrate LiquiColor recalled?
Three products requiring refrigerated storage conditions were incorrectly stored outside of approved temperature ranges. Using these temperature abused products may result in misdiagnosis of the patient.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot 193462
- Product
- B-Hydroxybutyrate LiquiColor, Item code CH2440058 - Product Usage: Product is used for the quantitative determination of B-Hydroxybutyrate in human…
- Recalling firm
- Cardinal Health Inc.
- Quantity
- N/A
- Distribution
- US Nationwide distribution NC, KS, and AL.
- Recall date
- May 27, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)