00597102010 Provisional CR ART SURF PROV 12/PURPLE 10 recall: Labeling Error
00597102010 Provisional CR ART SURF PROV 12/PURPLE 10, Rx, Sterile; 00597102012 Provisional CR ART SURF PROV 12/PURPLE 12, Rx, Sterile; 00597102014…
Why was 00597102010 Provisional CR ART SURF PROV 12/PURPLE 10 recalled?
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
all codes
- Product
- 00597102010 Provisional CR ART SURF PROV 12/PURPLE 10, Rx, Sterile; 00597102012 Provisional CR ART SURF PROV 12/PURPLE 12, Rx, Sterile; 00597102014…
- Recalling firm
- Zimmer, Inc.
- Quantity
- 192,355 all devices
- Distribution
- Worldwide Distribution-USA (nationwide) and the country of Japan.
- Recall date
- August 28, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)