FDAClass IIJuly 14, 2021

FREELITE Human Kappa Free Kit for use on the Roche recall

FREELITE Human Kappa Free Kit for use on the Roche cobas c Systems, Kappa Freie Leichtkette (Human) kit for Roche cobas c Systems, REF: LK016.CB, IVD…

Why was FREELITE Human Kappa Free Kit for use on the Roche recalled?

The high and low level "Quality Controls" demonstrate a positive bias of approximately 10% compared to assigned values.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers: 487719/ Exp: 31-Jul-2022, 480133/ Exp: 30-Sep-2022, Lot 457890/ Exp: 30-Sep-2022

Product
FREELITE Human Kappa Free Kit for use on the Roche cobas c Systems, Kappa Freie Leichtkette (Human) kit for Roche cobas c Systems, REF: LK016.CB, IVD…
Recalling firm
The Binding Site Group, Ltd.
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, GA, FL, MA, MO, MI, WA, NY, PA, MN, AL, MD, OK, NE, IN, ID, OH, TX, HI, DC and the countries of Canada, Australia, Austria, Belgium, Bulgaria, Canada, Cyprus, Czech Republic, Denmark, Estonia, France, Germany, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Slovenia, Spain, Switzerland, United Kingdom, Argentina, Egypt, GUADELOUPE, HONG KONG, India, Iran, Israel, Korea, Lebanon, Malaysia, Morocco, Oman, Singapore, Taiwan, Tunisia, Vietnam.
Recall date
July 14, 2021
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls