FDAClass IIJuly 16, 2014

The Hospira MedNet Medication Management Suite recall: Software Issue

The Hospira MedNet Medication Management Suite (MMS). Product Usage: The Hospira MedNet Medication Management Suite (MMS) is intended to facilitate…

Why was The Hospira MedNet Medication Management Suite recalled?

Hospira MedNet 5.5, 5.8.1, and 5.8.2 contains software defect where the dosing units of "nanog/kg/min" and "milliUnits/min" are not sent to a Plum A+ device (Version 13.40, 13.41, and 13.4.2) via Auto Programming when those dosing units are not properly established in the customized drug library. A dosing unit mismatch of "nanog/kg/min" or "milliUnits/min" on an unmatched medication will produce a

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

1) Software Version 5.5, List Number: 16037-04-13; 2) Software Version 5.8.1, List Number: 16037-04-31; 3) Software Version 5.8.2, List Number: 16037-04-33

Product
The Hospira MedNet Medication Management Suite (MMS). Product Usage: The Hospira MedNet Medication Management Suite (MMS) is intended to facilitate…
Recalling firm
Hospira Inc.
Quantity
13 units
Distribution
US Nationwide in the states of FL, LA, PA, WA, WY and the country of Hong Kong
Recall date
July 16, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls