The STERIS 4085 General Surgical Table Product Usage recall
The STERIS 4085 General Surgical Table Product Usage: is an electro-hydraulically operated surgical table designed to support all general surgical…
Why was The STERIS 4085 General Surgical Table Product Usage recalled?
STERIS engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All Model Numbers
- Product
- The STERIS 4085 General Surgical Table Product Usage: is an electro-hydraulically operated surgical table designed to support all general surgical…
- Recalling firm
- Steris Corporation
- Quantity
- 3689
- Distribution
- Worldwide Distribution - US (nationwide in the states of AL, AK, AZ, AR, CA,CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY in the countries of Canada, India, Vietnam, China, India, Korea, Mexico, costa Rica, Brazil, Taiwan, Australia, New Zealand, El Salvador, Indonesia, Hong Kong, Singapore, Nepal.
- Recall date
- July 16, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)