FDAClass IIJuly 2, 2025

CDI OneView System BPM Probe recall

CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit

Why was CDI OneView System BPM Probe recalled?

Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI 00886799001889 Lot Number/Serial Number Range: B0050231-B0050240 B0050221-B0050230 B0050211-B0050219 B0050201-B0050209 B0050191-B0050200

Product
CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit
Recalling firm
Terumo Cardiovascular Systems Corporation
Quantity
28 US; 7 OUS
Distribution
US Nationwide distribution in the states of California, Delaware, Hawaii, Iowa, Indiana, Michigan, North Carolina, New York, Ohio, and Washington. The country of New Zealand.
Recall date
July 2, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls