BIOGRAPH 40-3R to 64-3R Upgrade recall
BIOGRAPH 40-3R to 64-3R Upgrade, Material Number 10246390 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and…
Why was BIOGRAPH 40-3R to 64-3R Upgrade recalled?
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial Number 1001
- Product
- BIOGRAPH 40-3R to 64-3R Upgrade, Material Number 10246390 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and…
- Recalling firm
- Siemens Medical Solutions USA Inc.
- Quantity
- 1 unit
- Distribution
- Worldwide Distribution - US Nationwide
- Recall date
- June 29, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)