65597201131 Femur CR POR FEM Surf HDN Hatcp AML recall: Labeling Error
65597201131 Femur CR POR FEM SURF HDN HATCP AML Rx, Sterile; 65597201132 Femur CR POR FEM SURF HDN HATCP AMR Rx, Sterile; 65597201231 Femur CR POR FEM…
Why was 65597201131 Femur CR POR FEM Surf HDN Hatcp AML recalled?
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
all codes
- Product
- 65597201131 Femur CR POR FEM SURF HDN HATCP AML Rx, Sterile; 65597201132 Femur CR POR FEM SURF HDN HATCP AMR Rx, Sterile; 65597201231 Femur CR POR FEM…
- Recalling firm
- Zimmer, Inc.
- Quantity
- 192,355
- Distribution
- Worldwide Distribution-USA (nationwide) and the country of Japan.
- Recall date
- August 28, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)