FDAClass IIMay 20, 2020

Dover 100% Silicone Silver-Hydrogel Coated Premium recall

Dover 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage…

Why was Dover 100% Silicone Silver-Hydrogel Coated Premium recalled?

Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers: 1828206564, 1828813264 1829618264, 1829618364 1831001964, 1906506564 1907134164, 1909308864 1919705064, 1921106664 2000723764, 2003520164 2005617664

Product
Dover 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage…
Recalling firm
Covidien Llc
Quantity
46828 units
Distribution
US Nationwide distribution.
Recall date
May 20, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)