FDAClass IIJune 25, 2025

ICU Medical Plum Solos recall: Software Issue

ICU Medical Plum Solos, Item number 400011001; infusion pump

Why was ICU Medical Plum Solos recalled?

Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI/DI N/A, Serial/Lot Numbers: 60000001, 60000002, 60000003, 60000005, 60000006, 60000007, 60000008, 60000009, 60000010, 60000011, 60000014, 60000015, 60000016, 60000017, 60000019, 60000020, 60000021, 60000023, 60000024, 60000025, 60000026, 60000027, 60000028, 60000030, 60000032, 60000033, 60000035, 60000038, 60000039, 60000040, 60000041, 60000042, 60000044, 60000045, 60000046, 60000047, 60000048, 60000049, 60000050, 60000051, 60000052, 60000053, 60000054, 60000056, 60000057, 60000058, 60000059, 60000060, 60000061, 60000062, 60000063, 60000064, 60000065, 60000066, 60000067, 60000068, 60000069, 60000070, 60000071, 60000072, 60000073, 60000074, 60000075, 60000076, 60000077, 60000078, 60000079, 60000080, 60000081, 60000082, 60000083, 60000084, 60000085, 60000087, 60000088, 60000089, 60000090, 60000092, 60000093, 60000094, 60000095, 60000096, 60000097, 60000098, 60000099, 60000100, 60000101, 60000102, 60000104, 60000105, 60000107, 60000108, 60000109, 60000110, 60000112, 60000113, 60000114, 60000120, 60000130, 60000133, 60000134, 60000136, 60000139, 60000141, 60000143, 60000147, 60000150, 60000154, 60000159, 60000160, 60000162, 60000165, 60000167, 60000170, 60000171, 60000172, 60000173, 60000174, 60000176, 60000177, 60000178, 60000179, 60000180, 60000183, 60000187, 60000189, 60000190, 60000191, 60000192, 60000198, 60000199, 60000201, 60000203, 60000204, 60000205, 60000206, 60000207, 60000208, 60000209, 60000210, 60000212, 60000213, 60000214, 60000215, 60000216, 60000217, 60000223, 60000224, 60000225, 60000226, 60000227, 60000228, 60000229, 60000230, 60000234, 60000236, 60000237, 60000238, 60000239, 60000240, 60000241, 60000242, 60000244, 60000245, 60000246, 60000247, 60000249, 60000251, 60000252, 60000255, 60000257, 60000259, 60000263, 60000264, 60000265, 60000266, 60000268, 60000269, 60000271, 60000273, 60000274, 60000275, 60000276, 60000277, 60000279, 60000280, 60000281, 60000282, 60000283, 60000284, 60000285, 60000286, 60000288, 60000291, 60000292, 60000295, 60000296, 60000297, 60000298, 60000300, 60000155, 60000193, 60000197, 60000211, 60000243, 60000270, 60000272, 60000123, 60000168, 60000175, 60000195, 60000221, 60000256, 60000294.

Product
ICU Medical Plum Solos, Item number 400011001; infusion pump
Recalling firm
ICU Medical, Inc.
Quantity
214 units
Distribution
Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.
Recall date
June 25, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)