Olympus WM-NP2 or WM-DP2 Mobile Workstations with Nurse recall
Olympus WM-NP2 or WM-DP2 Mobile Workstations with Nurse Control Panel Arm, Model. No. MAJ-1664
Why was Olympus WM-NP2 or WM-DP2 Mobile Workstations with Nurse recalled?
WM-NP/DP2 workstations could become unstable and may fall over if the control panel arm is mounted with a video monitor weighing more than 19.8 pounds.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All serial numbers
- Product
- Olympus WM-NP2 or WM-DP2 Mobile Workstations with Nurse Control Panel Arm, Model. No. MAJ-1664
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- N/A
- Distribution
- Nationwide domestic distribution.
- Recall date
- July 24, 2019
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)