FDAClass IJuly 9, 2014

Langston Dual Lumen Catheters recall

Langston Dual Lumen Catheters, 6Fr, REF 5550 and 5540. The Vascular Solutions Langston Dual Lumen Pressure Monitoring catheter is used as a pressure…

Why was Langston Dual Lumen Catheters recalled?

The inner catheter of some Langston V2 Dual Lumen Catheters have separated from the device hub during contrast pressure injections. This may require an intervention to retrieve the separated piece from the patient to prevent injury.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers: 569436, 569437, 569962, , 569963, 569964, 569965, 570177, 570178, 570337, 570338, 570339, 570340, 570341, 570504, 570505, 570506, 570507, 570592, 570593, 570674, 570675, 570744, 570745, 570746, 570836, 570926, 570927, 570928, 570929, 571099, 571100, 571101, 571102, 571103, 571350, 571412, 571413, 571414, 571415, and 571727.

Product
Langston Dual Lumen Catheters, 6Fr, REF 5550 and 5540. The Vascular Solutions Langston Dual Lumen Pressure Monitoring catheter is used as a pressure…
Recalling firm
Vascular Solutions, Inc.
Quantity
8,580 total (US 5714)
Distribution
Worldwide Distribution - USA nationwide including DC and Puerto Rico, AUSTRIA, CANADA, GERMANY, ISRAEL, ITALY, NETHERLANDS, SPAIN, SWEDEN SWITZERLAND, and UNITED KINGDOM.
Recall date
July 9, 2014
Classification
Class I
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls