FDAClass IIJuly 24, 2019

Elekta Unity recall: Software Issue

Elekta Unity, Image-Guided Radiation Therapy System Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and…

Why was Elekta Unity recalled?

The QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI/GTIN: 5060081071321 Installed product: UNITY SYSTEM/13553-006/600003 UNITY SYSTEM/10817-011/600010 UNITY SYSTEM/12473-020/600020 UNITY SYSTEM/12274-017/600012 UNITY SYSTEM/10562-011/600016 UNITY SYSTEM/10719-T03001/600013 UNITY SYSTEM/30004070-001/600023 UNITY SYSTEM/10420-AVL-U/600008 UNITY SYSTEM/11014-UMCU-U/600007 UNITY SYSTEM/10157-006/600014 UNITY SYSTEM/11611-40/600011 UNITY SYSTEM/11489-21/600009

Product
Elekta Unity, Image-Guided Radiation Therapy System Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and…
Recalling firm
Elekta, Inc.
Quantity
12 units
Distribution
US Distribution to states of: TX & WI, and Internationally to: Canada, Denmark, Germany, Italy, Netherlands, Sweden, and the UK.
Recall date
July 24, 2019
Classification
Class II
Status
Terminated
Agency
FDA (device)