SteriPack Sterile Polyester Spun Swabs recall
SteriPack Sterile Polyester Spun Swabs (25 units/Pack)- Intended for sample collection in the intended nasal area. Catalog Number: 60566RevB
Why was SteriPack Sterile Polyester Spun Swabs recalled?
Fails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention to remove part of the swab
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
GTIN-DI: 00850027193205 Pack Lot Number : 85438; Individual Swab Lot Number: 86445 Exp Date: 28-Oct-2023
- Product
- SteriPack Sterile Polyester Spun Swabs (25 units/Pack)- Intended for sample collection in the intended nasal area. Catalog Number: 60566RevB
- Recalling firm
- LumiraDx
- Quantity
- 1003 kits
- Distribution
- US Nationwide distribution.
- Recall date
- June 21, 2023
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)