Vaxcel PICC with PASV MST-60 Kit under the following recall
Vaxcel PICC with PASV MST-60 Kit under the following labels: 1) 3F, UPN H965454360, Rx ONLY, 2) 4F, UPN M001454590, Rx ONLY, 3) 5F, UPN's M001454640…
Why was Vaxcel PICC with PASV MST-60 Kit under the following recalled?
PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Batch/Lots: 1) 4724945, 4731137, 4738296, 4895232, 4912062, 4933353, 2) 4731139, 4738302, 4748674, 4776152, 3) 4721239, 4784467, 4810672, 4812161, 4821239, 4821511, 4824106, 4830612, 4832722, 4836599, 4724949, 4724950, 4738298, 4738299, 4773297, 4773298, 4786764, 4823438, 4829478, 4832723,4845209, 4724926 & 4734253 with Use By Date Range 2016-03-31 to 2017-09-30.
- Product
- Vaxcel PICC with PASV MST-60 Kit under the following labels: 1) 3F, UPN H965454360, Rx ONLY, 2) 4F, UPN M001454590, Rx ONLY, 3) 5F, UPN's M001454640…
- Recalling firm
- Navilyst Medical, Inc., an AngioDyamics Company
- Quantity
- 5,566 units
- Distribution
- Worldwide Distribution US (Nationwide) Internationally to AU, GR, SE, SA, IN, CA, GB, BR, BE, NL, ES, KR, HK, AE, IT and ES.
- Recall date
- June 22, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)